DCA Vantage Service Manual Version 1.0 052907 DCR Optimized
DCA Vantage Service Manual Version 1.0 052907 DCR Optimized
1 Overview
Intended Use
The DCA Vantage is a semi-automated, benchtop system. It is designed to quantitatively
measure the percent of Hemoglobin A1c in blood and low concentrations of albumin in urine
(microalbuminaria), creatinine in urine, and the albumin/creatinine ratio in urine.
The DCA Vantage system is intended for professional use in a physician’s office or hospital
laboratory. Tests performed using the DCA Vantage system are intended for in vitro
diagnostic use. As with all diagnostic tests, do not base a definitive diagnosis on the results of
a single test. A physician should make a diagnosis after all clinical and laboratory findings
are evaluated.
Hardware Overview
User Interface
By default, interaction with the DCA Vantage system is via an integrated touchscreen.
Messages, options, and requests for information display, and responses are made by selecting
the appropriate key symbol on the screen.
CAUTION: Do not use anything hard or pointed on the touchscreen. It may damage
the screen.
You can also use a handheld barcode reader to interact with the system.
The DCA Vantage system is divided into the following 3 modules:
• Test Sequences
• Recall
• System
The barcode reader is used to calibrate the system and scan the reagent cartridges and control
cards. The reagent cartridges are placed in the reagent cartridge compartment where the tests
are run, and when testing is complete, an internal thermal printer prints the test results.
2 Printer
3 Display Screen
Memory
The operating system, the system software, the test data, and the system settings for the DCA
Vantage system are stored in an onboard (non-removable) flash memory. You can insert a
USB Flash memory drive into the system’s USB port during the Backup sequence. You can
then insert the USB memory drive into another DCA Vantage system and restore the data to
the new system.
You can transfer patient data from one DCA Vantage system to another using the software
backup and restore sequences in system settings, refer to System Configuration on page 9.
1-4 Overview
You cannot import data from a spreadsheet into a DCA Vantage system.
1 Connections
2 Power Switch
3 Power Cord
Software Overview
The DCA Vantage system provides an easy-to-navigate and intuitive user interface. When
the system is not in use for more than 30 minutes, the Power Save Mode automatically turns
on. Touch any ___location on the screen to resume operation.
Screen Components
The DCA Vantage system software screens use a consistent layout.
1 Screen Title
Overview 1-5
4 Title Bar
5 Content Area
Some options require that you enter information. If selected, a numeric keypad or an
alphanumeric keypad displays. An onboard barcode scanner is included and reads only
barcodes of DCA cartridges and calibration cards.
The DCA Vantage system does not support the connection of an external computer keyboard
or a mouse to enter data.
1-6 Overview
Spare parts 2-1
This section will provide pictures and part numbers for the spare part available for the DCA
Vantage instrument
3
1
Figure 1
Find
Description BAN
Number
1 Reagent compartment Door 06498697
Top Housing molding with window
2 06498700
Note – This is only available as special order
3 Printer Cover 06498689
Spare parts 2-3
4
2
Figure 2
Find
Description BAN
Number
1 Graphical Display with Touchscreen 06498549
2 Door Switch Assembly 06498719
3 Barcode Readhead 06498573
4 DCA Main PCB assembly with bracket 06498522
5 Display Cable 06498557
6 Printer 06498565
Spare parts 2-4
2
5
Figure 3
Find
Description BAN
Number
1 Motor Assembly 06498581
2 Cartridge Holder Assembly 06498654
3 Optical Switch – Home Sensor 06498603
4 Locking Yoke 06498611
5 Preamp Assembly 06498670
Spare parts 2-5
Figure 4
Find
Description BAN
Number
1 Source Optics Assembly Complete 06498638
2 Lamp 06498646
3 Filter - Optical 06498662
Spare parts 2-6
7
2
4
5 Figure 5
Find
Description BAN
Number
1 Pad Rubber with adhesive 06498395
2 Cover Printer Paper 06498379
3 Fan 06498409
4 Speaker 06498983
5 Wire Harness DC Power 06498441
6 Power Supply Assembly 06498433
Fuse (not shown in figure 2 - required) 06498506
7 Fan Filter Bezel 06498417
8 Filter Material 06498425
Spare parts 2-7
Figure 5
Find
Description BAN
Number
1 DCA Assembly Inverter PCB 06498530
Spare parts 2-8
The following items are not shown in any figures in this section.
Description BAN
Cord Set UK – 3 Pin 5 Amp Plug 06498298
USA Cord Set 115 Volt 06498301
Cord Set Europe Moulded Plug 06498352
Printer Paper (5 pack) 06498379
Printer Paper Self Adhesive Stock (pack of 5) 06352780
Optical Test Cartridge 06489221
DCA Cleaning Kit (10) 06488209
Barcode Scanner Kit 06502880
Bar Code Scanner Cable 06498247
Bar Code Scanner Stand 06761052
Collimating Tube (Pak of 2) 06498360
Replacement 3-1
3 Replacement Section
In this section step by step procedures will be provided for the replacement of the serviceable
items in the DCA Vantage instrument.
A B
C
Figure 3-1
Replacement 3-4
B
C
D A
Figure 3-2
5. Referring to figure 3-3; lift up the plastic shield and remove heater cable screw “A”,
motor cable connector “B”.
6. Again referring to figure 3-3 disconnect; heater cable from connector “C”, home sensor
cable connector “D”, lamp leads from screw terminals “E” and preamp cable from the
zero insert force (ZIF) locking connector “F”. The motor assembly is now completely
disconnected from the main PCB.
7. The middle case can now be lifted up and laid to the left of the lower case, see figure 3-4.
8. Wire harness, “F” in figure 3-4 connects the middle case to the lower case assembly.
Disconnecting connectors identified in figure 3-4 as: A, B, C and E will free the middle
case from wire harness “F”.
Replacement 3-5
B D
Figure 3-3 G
A B C
D E
Figure 3-4
Replacement 3-6
Figure 3-5 D
Figure 3-6
B
A C D
Figure 3-7
C D
A B
E
F Figure 3-8
6. Reinstall the four screws “C”, “D”, “E” and “F” in figure 3-8 that hold the display
assembly to the middle case assembly.
7. Referring to figure 3-8 reconnect the touch screen cable “A” and display data cable “B”
8. Guide the display backlight cable through the middle case.
9. Reconnect the display back light cable to connector “A” in figure 3-7 on the power
inverter PCB
Replacement 3-10
3.6.2 Procedure
1. Remove the upper case (refer to section 3.1.2).
2. Referring to figure 3-9, disconnect ribbon cables “A” and “B”.
3. Remove screw “C” which holds the printer to the case.
4. To remove the printer slide it towards the main PCB this will disengage the plastic
fingers “D” and allow the printer to be lifted out.
B
A
C
Figure 3-9
3.7.2 Procedure:
1. Remove the upper case (refer to section 3.1.2).
2. Referring to figure 3-10, the barcode readhead “A” is connected to the main PCB by a
cable “C” and two screws “B”.
A
C
D
B
Figure 3-10
A
A
A
B
B
Figure 3-11
A
B
Rear Screws
C
A
Front Screws
Figure 3-12
A
C
Figure 3-13
5. To remove the lamp loosen screw “B”, this will permit the lamp to be removed from the
optics assembly.
2. Reinstall the optics assembly into the Motor mount assembly. Refer to figure 3-14 “A”
for optical filter orientation, the “Gold” colored side of the filter should face out. The
other side of the filter rests against two foam washes “B” on the optics housing. When
installing the assembly the optical filter will be aligned in the Motor assembly by three
guides “C”.
3. Once the optical assembly is in place in the Motor mount assembly secure in place using
the screw “A” in figure 3-13.
4. Reinstall the motor mount assembly (refer to section 3.9.3).
5. Reinstall the middle case (refer to section 3.2.3).
6. Reinstall the upper case (refer to section 3.1.3).
A
B
Figure 3-14
6. With the preamp PCB assembly removed the collimating tubes “D” can be accessed and
should be blown out with clean air to clear the tubes of dust before the preamp PCB is
reinstalled.
A
C
Figure 3-15
C
C
C C
B
Figure 3-16
Replacement 3-18
Figure 3-17
Replacement 3-19
Figure 3-18
Troubleshooting 5-1
5 Troubleshooting
General Information
If an operational or system problem occurs, an error code may display on the system screen
with an explanation of the problem. This section of the guide lists the various errors and
messages, along with a description and corrective actions. If the problem persists, record the
error code and contact your local technical service provider for assistance.
If you think DCA cartridges are causing the problem, consult the product insert that comes
with the cartridges for troubleshooting information.
If you turn the system off, you must retest all samples that were in process when the error
occurred.
With some errors, the system continues to run while the error displays.
CAUTION: Wait at least 15 minutes after the reagent cartridge is removed from the
refrigerator to start a test.
CAUTION: Verify that the capillary is completely filled with no air bubbles.
Troubleshooting 5-2
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
Door Open Error The door was open Any sample test door opened or failed Close door 1) Check door sensor
during a test (HbA1c, door sensor rerun test and verify cable is
Microalbumin/Creati connected.
nine) or Optical 2) Replace door sensor.
Filter Test (S1, S2 3) Replace Main PCB
or S3), or during
Motor Homing
operation.
E99 – Internal This is a "catch all" Any time Software or Hardware Call Service Check hardware and
System Error error for internal Bug notify Global Service
system failures and Support
that should never
happen. If this
error is posted, it
usually indicates a
bug in the system.
E3 The DCA analyzer The DCA analyzer The memory device 1) Insert the 1) USB Device
is unable to access is unable to access was disconnected or same USB not seated correctly
the memory device the memory device malfunctioned before memory remove and reseat.
containing a containing a the software update device that 2) Defective
software update software update was completed. was used for USB memory device.
(usually a USB (usually a USB flash the update 3) Defective
flash drive). drive). and press PCB Replace.
"OK" in the
dialog box that
is displaying
the "E3" error.
The system
will then
reboot itself.
2) Customer
contacts
Service
E10 Motor The Home/Index During the Motor Failure of Home/Index Re boot 1) Re-boot instrument
position sensor sensor failed to Homing operation sensor, Sensor circuit, instrument 2) Test optical sensor
error detect light-to-dark following power-up motor, or motor drive using hardware tests.
or dark-to-light or during the Motor circuit; or obstruction Contact 3) Check sensor
transitions during a Homing operation to cartridge holder Service alignment.
Motor Homing performed when a rotation 4) Blow out sensor
operation. OR The test is aborted. possible dust
offset between the 5) Sensor bad
Home position Replace
determined from 6) PCB failure replace
the Home/Index PCB
senor and the
position for the
optimum optical
signal is greater
than +/- 4 half-
steps.
E12 Motor The wrong number During any motor Failure of Home/Index Re-boot 1) Re-boot instrument
Position error of Home/Index rotation that occurs sensor, Sensor circuit, instrument 2) Test optical sensor
sensor transitions during a sample motor, or motor drive using hardware tests.
were detected test, optics test, or circuit; or obstruction Contact 3) Check sensor
during a motor Motor position test to cartridge holder service alignment.
rotation operation. in service. rotation 4) Blow out sensor
possible dust
5) Sensor bad
Replace.
6) PCB failure replace
PCB
Troubleshooting 5-3
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
E20 Dark offset A "Dark" reading During any sample Electronic failure Contact 1) Check that Preamp
measurement out (mean of 16) for test (HbA1c, service connector is correctly
of range sample or Microalbumin/Creati seated.
reference channel nine), or Optical 2) Possible pre-amp
is outside the Test (S1, S2, or S3). PCB failure, replace
acceptance range. PCB
(see 4.2.3.1) 3) Possible main PCB
failure, replace main
PCB
E21 Lamp failure An "Air" reading During Power On Incorrect position of 1) Re-boot 1) Blow out cartridge
(mean of 16) for Initialization or any cartridge holder; or instrument. area.
sample or sample test (HbA1c, failure of lamp, lamp 2) Clean 2) Verify cartridge
reference channel Microalbumin/Creati drive circuit, cartridge holder in correct
is too low. (see nine), Optical Filter photodiode/preamp holder area, ___location.
4.2.3.1) Test (S1, S2 or S3). circuit, or analog data dust blocking 3) Check for lamp
acquisition reference function. If not
electronics. channel light functional confirm
path. (use voltage. If voltage ok
compressed replace lamp, if voltage
air) not present replace
Main PCB.
4) If lamp functional
clean optics and
collimation tubes
E22 Optical A Sample or During Power On Failure of 1) Re-boot 1) Blow out cartridge
Reading Out of Reference channel initialization or any photodiode/preamp instrument. area.
Range reading (mean of sample test (HbA1c, detector circuit, lamp 2) Clean 2) Verify cartridge
16) taken during Microalbumin/Creati drive circuit or optical cartridge holder in correct
an Air nine), Optical Filter system holder area, ___location.
measurement is Test (S1, S2, or S3). dust blocking 3) Check for lamp
too high (greater reference function. If not
than the reliable channel light functional confirm
measurement path. (use voltage. If voltage ok
range). (see compressed replace lamp, if voltage
4.2.3.1) air) not present replace
Main PCB.
4) If lamp functional
clean and / or replace
optics and collimation
tubes
5) Replace Pre Amp
PCB
E23 Excessive The standard During Power On Low lamp intensity, or 1) Re-boot 1) Blow out cartridge
Noise in Sample deviation of the 16 Initialization, any lamp instrument. area.
Channel Reading readings on the sample test, or any alignment/rotation, or 2) Clean 2) Verify cartridge
Sample channel optical test lamp intensity cartridge holder in correct
for Dark or Air variation, or electronic holder area, ___location.
reading is too defect dust blocking Check for lamp function.
large. (NOTE: reference 3) If not functional
This test cannot be channel light confirm voltage. If
performed until the path. (use voltage ok replace lamp,
first air compressed if voltage not present
measurement is air) replace Main PCB.
taken.) (see 4) If lamp functional
4.2.3.2) clean and / or replace
optics and collimation
tubes
5) Replace Preamp
PCB
E24 Excessive The standard During Power On Low lamp intensity, or 1) Re-boot 1) Blow out cartridge
Noise in deviation of the 16 Initialization or any lamp instrument. area.
Reference readings on the sample test (HbA1c, alignment/rotation, or 2) Clean 2) Verify cartridge
Channel Reading Reference channel Microalbumin/Creati variation in lamp cartridge holder in correct
for Dark, Air, or nine), Optical Filter intensity, or electronic holder area, ___location.
Sample reading is Test (S1, S2 or S3). defect dust blocking 3) Check for lamp
Troubleshooting 5-4
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
too large. (NOTE: reference function.
This test cannot be channel light 4) If not functional
performed until the path. (use confirm voltage. If
first air compressed voltage ok replace lamp,
measurement is air) if voltage not present
taken.) (see replace Main PCB.
4.2.3.2) 5) If lamp functional
clean and / or replace
optics and collimation
tubes
6) Replace Preamp
PCB
E26 Excessive The standard Any sample test Low lamp intensity, or 1) Re-boot 1) Blow out cartridge
Noise in Sample deviation of the 16 (HbA1c, lamp instrument. area.
Channel During readings on the Microalbumin/Creati alignment/rotation, or 2) Clean 2) Verify cartridge
Sample Reading Sample channel nine) or Optical lamp intensity cartridge holder in correct
during Filter Test (S1, S2 variation, or holder area, ___location.
a reading taken in or S3). electronic defect dust blocking 3) Check for lamp
Sample read reference function.
position. (NOTE: channel light 4) If not functional
This test cannot be path. (use confirm voltage. If
performed until the compressed voltage ok replace lamp,
first air air) if voltage not present
measurement is replace Main PCB.
taken.)(see 5) If lamp functional
4.2.3.2) clean and / or replace
optics and collimation
tubes
6) Replace Preamp
PCB
E27 Excessive The change in During any sample Dust contamination in 1) Re-boot 1) Blow out cartridge
Lamp Drift mean signal test (HbA1c, the optical path, or instrument. area.
between Microalbumin/Creati lamp filament 2) Clean 2) Verify cartridge
successive Air nine), or Optical orientation, or lamp cartridge holder in correct
readings is too Filter Test (S1, S2 variation, or electronic holder area, ___location.
large on either the or S3). defect. dust blocking 3) Check for lamp
Sample channel or reference function.
the Reference channel light 4) If not functional
channel. (NOTE: path. (use confirm voltage. If
This test cannot be compressed voltage ok replace lamp,
performed until the air) if voltage not present
second air replace Main PCB.
measurement is 5) If lamp functional
taken.) (see clean and / or replace
4.2.3.3) optics and collimation
tubes
6) Replace Preamp
PCB
E30 Cartridge The temperature Continuously during Major defect in heater 1) Re-boot 1) Check to make
Temperature measured by one system idle time and plate/thermistor instrument. sure that heater cable is
Extremely Low of the cartridge during Sample tests. assembly, connector, fully inserted into the
holder thermistors or electrical circuit; or Main PCB
is <= 2ºC. ambient temperature 2) Replace the
<4ºC (far out of Cartridge heater
operating range) assembly.
3) Replace the Main
PCB.
E31 Cartridge The temperature Continuously while Cold cartridge; or 1) Make 1) Check to make
Temperature Out measured by one the system is idle failed heater sure that sure that heater cable is
of Range Low of the cartridge and during Optical plate/thermistor reagent fully inserted into the
holder thermistors Test. Also during assembly, connector cartridge Main PCB
is <= (Tsetpoint - sample tests as or electrical circuit handling 2) Replace the
12.9)ºC while the follows .HbA1c: procedures in Cartridge heater
system is idle, or from the start of the reagent assembly.
during an Optical test until 3 minutes product insert 3) Replace the Main
Troubleshooting 5-5
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
Test, the first 3 after the start have been PCB.
minutes of an Microalbumin followed.
HbA1c test, or the /Creatinine: from the 2) Reboot
first 5 minutes or a start of the test until system and
Microalbumin/ 5 minutes after the repeat test.
Creatinine test. start
E32 Cartridge The temperature HbA1c: 3 minutes Cold cartridge; or 1) Make 1) Check to make
Temperature measured by one after start of test failed heater sure that sure that heater cable is
Below Limit of the cartridge until the end plate/thermistor reagent fully inserted into the
During Critical holder thermistors Microalbumin / assembly, connector cartridge Main PCB
Measurement is <= (Tsetpoint - Creatinine: 5 or electrical circuit handling 2) Replace the
0.9)ºC during minutes after start of procedures in Cartridge heater
critical test until the end reagent assembly.
measurements of a product insert 3) Replace the Main
sample test have been PCB.
sequence. followed.
2) Reboot
system and
repeat test.
E33 Cartridge The temperature HbA1c: 3 minutes Cold cartridge; or 1) Reboot 1) Check to make
Temperature measured by one after start of test failed heater system and sure that heater cable is
Above Limit of the cartridge until the end plate/thermistor repeat test. fully inserted into the
During Critical holder thermistors assembly, connector Main PCB
Measurement is >= (Tsetpoint + Microalbu or electrical circuit 2) Replace the
1.1)ºC during min / Creatinine: 5 Cartridge heater
critical minutes after start of assembly.
Measurements of a test until 3) Replace the Main
sample test PCB.
sequence. the end
E34 Cartridge The temperature Continuously while Cold cartridge; or 1) Reboot 1) Check to make
Temperature Out measured by one the system is idle failed heater system and sure that heater cable is
of Range High of the cartridge and during Optical plate/thermistor repeat test. fully inserted into the
holder thermistors Test. Also during assembly, connector Main PCB
is >= (Tsetpoint + sample tests as or electrical circuit 2) Replace the
2.1)ºC follows..HbA1c: Cartridge heater
from the start of the assembly.
test until 3 minutes 3) Replace the Main
after the start PCB.
Microalbumin
/Creatinine: from the
start of the test until
5 minutes after the
start
E35 Cartridge The temperature Continuously during Major defect in heater 1) Reboot 1) Check to make
Temperature measured by one system idle time and plate/thermistor system and sure that heater cable is
Extremely High of the cartridge during Sample tests. assembly, connector, repeat test. fully inserted into the
holder thermistors or electrical circuit; Main PCB
is >= 50ºC. ambient temperature 2) Replace the
too high (far out of Cartridge heater
operating range) assembly.
3) Replace the Main
PCB.
E36 Temperature 10 minutes after Continuously Failure of heater 1) Reboot 1) Check to make
After 10 Minutes power-on, the plate/thermistor system and sure that heater cable is
Still Out of Range cartridge assembly, connector, repeat test. fully inserted into the
temperature has heater driver, or Main PCB
not reached the electronics 2) Replace the
operating range of Cartridge heater
(Tsetpoint - 13.2)ºC < assembly.
T < (Tsetpoint + 3) Replace the Main
1.8)ºC, or the PCB.
cartridge
temperature
Troubleshooting 5-6
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
remains outside
the same operating
range
for more than 10
minutes.
E37 Internal The temperature Continuously during Damaged ambient 1) Reboot 1) Reset the ambient
Instrument measured by the system idle time and temperature system and temperature sensor.
Temperature ambient during Sample tests thermistor or circuit repeat test. 2) Replace the Main
Extremely Low temperature failure PCB.
thermistor is <=
2ºC
E38 Internal The temperature Continuously during Damaged ambient 1) Reboot 1) Reset the ambient
Instrument measured by the system idle time and thermistor or circuit system and temperature sensor.
Temperature ambient during Sample tests. failure repeat test. 2) Replace the Main
Extremely High temperature PCB.
thermistor is >=
50ºC
E40 Incorrect The number of Any time a bar code Incorrect bar code; or 1) Rescan 1) Bar code reader
Number of Bar digits read by the is read partial bar code scan; the bar code head dirty clean with air.
Code Digits bar code reader or failure of bar code in different 2) Bar code reader
does not match reader. direction window dirty / scratched
what is expected. 2) Try a replace upper case.
different 3) Bar code reader
barcode head damaged replace
source reader head.
cartridge or 4) Main PCB bad
Calibration replace.
card
3) Bar code
window is dirty
–clean
E41 Incorrect Bar The bar code has Any time a bar code Incorrect bar code 1) Rescan 1) Bar code reader
Code Data the correct number scanning is allowed data or bar code read the bar code head dirty clean with air.
of digits, but the error in different 2) Bar code reader
character does not direction window dirty / scratched
match the 2) Try a replace upper case.
expected pattern different 3) Bar code reader
barcode head damaged replace
source reader head.
cartridge or 4) Main PCB bad
Calibration replace.
card
3) Bar code
window is dirty
–clean
E50 Serial Port An attempt to Any time the Bad/missing 1) Reseat 1) Test serial port
Communication communicate via analyzer attempts to connection or printer cable using loop back
failure the RS232 serial communicate with electronic failure on Vantage connector. Replace
port has failed an external and at main PCB if necessary
computer / LIS destination
2) Check
that the
RS232 serial
port on the
Vantage is
configured to
meet the
requirements
of the host
device.
3) Verify
that serial port
works with a
Troubleshooting 5-7
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
loop back
connector or
PC using
hyper
terminal.
4) Return
instrument for
service
E51 Serial Port Communication Whenever the Bad/missing same as 1) Test Ethernet port
Communication has not succeeded analyzer attempts to connection or above Replace main PCB if
Error even after the communicate with electronic failure necessary same as
allowed retries are an external above
exhausted. computer / LIS via
the RS232 Serial
Port and there is
already an E50
error.
E55 Ethernet Port An attempt to Any time the Bad/missing 1) Reseat same as above
Communication communicate via analyzer attempts to connection or Ethernet cable
Failure the Ethernet serial communicate with electronic failure on Vantage
port has failed. and external and at
computer / LIS via destination
the Ethernet port 2) Check
that the
Ethernet port
on the
Vantage is
configured to
meet the
requirements
of the host
device.
3) Return
instrument for
service
E56 Ethernet Port Communication Whenever the Bad/missing same as same as above
Communication has not succeeded analyzer attempts to connection or above
Error even after the communicate with electronic failure
allowed retries are an external
exhausted computer / LIS via
the Ethernet
connection and
there is already a
E55 error
E60 Database The database is During database Corruption of the 1) Re-boot 1) Reboot the system
Failure not functioning - compaction or when database the system 2) Re-initialize
cannot read or opening the database with another
write. database application
E61 Database A database read An error was Corruption of the 1) Re-boot 1) Reboot the system
Read Failure operation has returned from a database the system 2) Re-initialize
failed resulting in database write database with another
no data or operation. application
corrupted data
being returned
D62 Database An error was An error was Corruption of the 1) Re-boot 1) Reboot the system
Write Error returned from a returned from a database the system 2) Re-initialize
database write database write database with another
operation. operation. application
E63 Database The database has Immediately after The database has 1) Delete 1) Reboot the system
Full completely filled its saving the last reached its allotted some records 2) Re-initialize
allotted capacity sample test result limit 2) Re-boot database with another
for storing sample has filled the system application
test results. database.
Troubleshooting 5-8
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
E101 HbA1c The mean During HbA1c test No cartridge present; 1) Make 1) Blow out cartridge
Buffer Mean transmittance of sequence or optical problem; or sure cartridge area.
Absorbance Too the 10 buffer electronic drift is fully 2) Verify cartridge
Low readings inserted and holder integrity and
(buf1...buf10) is locked in motor positions cartridge
too high. place. in correct ___location.
2) Review 3) Check for lamp
cartridge function and drift.
handling Replace lamp if
instructions necessary.
3) Inspect 4) If lamp functional
cartridge clean and / or replace
window for optics and collimation
scratches or tubes
finger prints 5) Replace Preamp
4) Run PCB
optical test 6) Replace Main
cartridge. PCB.
5) Clean
cartridge
holder area,
dust blocking
reference
channel light
path. (use
compressed
air)
6) Rerun
test
E102 HbA1c The mean During HbA1c test Condensation on 1) Make 1) Blow out cartridge
Buffer Mean transmittance of sequence cartridge optical sure cartridge area.
Absorbance Too the 10 buffer window; or cartridge is fully 2) Verify cartridge
High readings not located correctly inserted and holder integrity and
(buf1...buf10) is in holder; or defect on locked in motor positions cartridge
too low. optical cartridge place. in correct ___location.
optical window; or 2) Review 3) Check for lamp
buffer tab not pulled; cartridge function and drift.
or cartridge use life handling Replace lamp if
expired; or optical instructions necessary.
problem; or electronic 3) Inspect 4) If lamp functional
drift cartridge clean and / or replace
window for optics and collimation
scratches or tubes
finger prints 5) Replace Preamp
4) Run PCB
optical test 6) Replace Main
cartridge. PCB.
5) Clean
cartridge
holder area,
dust blocking
reference
channel light
path. (use
compressed
air)
6) Rerun
test
E103 Buffer The coefficient of During HbA1c test Condensation of 1) Make 1) Blow out cartridge
Coefficient of variation (%CV) of sequence cartridge optical sure cartridge area.
Variation Too the 10 window; or particulate is fully 2) Verify cartridge
High transmittance contamination in inserted and holder integrity and
readings the buffer cartridge; or cartridge locked in motor positions cartridge
solution in the use life expired; or place. in correct ___location.
reagent cartridge is high "short term" 2) Review 3) Check for lamp
too high. instrument drift cartridge function and drift.
Troubleshooting 5-9
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
handling Replace lamp if
instructions necessary.
3) Inspect 4) If lamp functional
cartridge clean and / or replace
window for optics and collimation
scratches or tubes
finger prints 5) Replace Preamp
4) Run PCB
optical test 6) Replace Main
cartridge. PCB.
5) Clean
cartridge
holder area,
dust blocking
reference
channel light
path. (use
compressed
air)
6) Rerun
test
E104 Blood The transmittance During HbA1c test No sample; low total 1) Make 1) Blow out cartridge
Pickup of the blood pickup sequence hemoglobin; or sure cartridge area.
Absorbance Too reading is too high. capillary not is fully 2) Verify cartridge
Low completely filled or inserted and holder integrity and
bubble locked in motor positions cartridge
place. in correct ___location.
2) Review 3) Check for lamp
cartridge function and drift.
handling Replace lamp if
instructions necessary.
and sample 4) If lamp functional
device filling clean and / or replace
procedures. optics and collimation
3) Inspect tubes
cartridge 5) Run performance
window for test.
scratches or
finger prints
4) Rerun
test
E105 Blood The transmittance During HbA1c test Too much blood (drop 1) Make 1) Blow out cartridge
Pickup of the blood pickup sequence of blood on outer sure cartridge area.
Absorbance Too reading is too low. surface of capillary); is fully 2) Verify cartridge
High or total inserted and holder integrity and
hemoglobin too high; locked in motor positions cartridge
or cartridge not place. in correct ___location.
allowed to warm up; 2) Review 3) Check for lamp
blood not lysing; or cartridge function and drift.
cartridge use life handling Replace lamp if
expired; or blood instructions necessary.
sample cell and sample 4) If lamp functional
irregularity or device filling clean and / or replace
abnormal hemoglobin procedures. optics and collimation
3) Inspect tubes
cartridge 5) Run performance
window for test.
scratches or
finger prints
4) Rerun
test
E106 Hemoglobin The mean During HbA1c test Total hemoglobin < 7 1) Make 1) Blow out cartridge
Absorbance Too transmittance of sequence g/dL; or under-filled sure cartridge area.
Low the hemoglobin sample capillary; or is fully 2) Verify cartridge
readings is too bubble in sample inserted and holder integrity and
Troubleshooting 5-10
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
high locked in motor positions cartridge
place. in correct ___location.
2) Review 3) Check for lamp
cartridge function and drift.
handling Replace lamp if
instructions necessary.
and sample 4) If lamp functional
device filling clean and / or replace
procedures. optics and collimation
3) Inspect tubes
cartridge 5) Run performance
window for test.
scratches or
finger prints
4) Rerun
test
E107 Hemoglobin The mean During HbA1c test Total hemoglobin > 1) Make 1) Blow out cartridge
Absorbance Too transmittance of sequence 24 g/dL; excess blood sure cartridge area.
High the hemoglobin on sides of is fully 2) Verify cartridge
readings is too low capillary(not wiped inserted and holder integrity and
off); or wicking of locked in motor positions cartridge
blood between the place. in correct ___location.
capillary and capillary 2) Review 3) Check for lamp
holder; or blood not cartridge function and drift.
lysing (cold cartridge); handling Replace lamp if
blood cell irregularity; instructions necessary.
or abnormal and sample 4) If lamp functional
hemoglobin device filling clean and / or replace
procedures. optics and collimation
3) Inspect tubes
cartridge 5)
window for
scratches or
finger prints
4) Rerun
test
E108 %CV of The %CV of the During HbA1c test Particulate 1) Review 1) Blow out cartridge
Hemoglobin transmittance of sequence contamination in cartridge area.
Readings Too the hemoglobin cartridge; or high storage and 2) Verify cartridge
Large readings is too "short term" handling holder integrity and
large. instrument drift instructions. motor positions cartridge
2) Inspect in correct ___location.
cartridge 3) Check for lamp
window for function and drift.
scratches and Replace lamp if
finger prints necessary.
3) Run 4) If lamp functional
optical test clean and / or replace
cartridge. optics and collimation
4) Rerun tubes
test. 5) Run performance
test.
E109 Individual The absorbance of During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Agglutination one of the sequence excessive cartridge problem
Reading Too Low agglutinations is temperature and/or storage and
too low humidity; or low or no handling
agglutination instructions.
2) Inspect
cartridge
window for
scratches and
finger prints
3) Rerun
test.
E110 Individual The absorbance of During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Troubleshooting 5-11
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
Agglutination one of the sequence excessive cartridge problem
Reading Too High agglutinations is temperature and/or storage and
too high. humidity; or high handling
agglutination instructions.
2) Inspect
cartridge
window for
scratches and
finger prints
3) Rerun
test.
E111 Delta in The delta in During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Agglutination absorbance sequence excessive cartridge problem
Absorbance Too between the first temperature and/or storage and
Low and last humidity; or low or no handling
agglutination agglutination instructions.
reading is too low. 2) Inspect
cartridge
window for
scratches and
finger prints
3) Rerun
test.
E112 Delta in The delta in During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Agglutination absorbance sequence excessive cartridge problem
Absorbance Too between the first temperature and/or storage and
High and last humidity handling
agglutination instructions.
reading is too high. 2) Inspect
cartridge
window for
scratches and
finger prints
3) Rerun
test.
E113 Third Order The A1 parameter During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Polynomial Fit from the third order sequence excessive cartridge problem
Parameter A1 Too polynomial fit is an temperature and/or storage and
Negative unexpectedly large humidity; or handling
negative number. particulate instructions.
contamination in 2) Inspect
cartridge cartridge
window for
scratches and
finger prints
3) Rerun
test.
E114 Third Order The A2 parameter During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Polynomial Fit from the third order sequence excessive cartridge problem
Parameter A2 Too polynomial fit is too temperature and/or storage and
Large large. humidity; or handling
particulate instructions.
contamination in 2) Inspect
cartridge cartridge
window for
scratches and
finger prints
3) Rerun
test.
E115 Large The difference During HbA1c test Cartridge exposed to 1) Review Most likely a reagent
Change in between the blood sequence excessive cartridge problem
Hemoglobin pickup absorbance temperature and/or storage and
Absorbance and the mean humidity; or handling
Reading hemoglobin particulate instructions.
absorbance is too contamination in 2) Inspect
Troubleshooting 5-12
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
large. cartridge cartridge
window for
scratches and
finger prints
3) Rerun
test.
E116 Mean The Mean Squared During HbA1c test Cartridge defect; or 1) Review Most likely a reagent
Squared Error of Error calculated sequence high sample cartridge problem
Agglutination after the third order absorbance storage and
Readings Too polynomial fit of handling
Large the agglutination instructions.
readings is too 2) Inspect
large. cartridge
window for
scratches and
finger prints
3) Rerun
test.
E301 The mean During No cartridge present; 1) Make 1) Blow out cartridge
Microalbumin / transmittance of Microalbumin / or optical problem; sure cartridge area.
Creatinine Buffer the 10 buffer Creatinine test electronic drift is fully 2) Verify cartridge
Mean Absorbance readings sequence inserted and holder integrity and
Too Low (buf1...buf10) is locked in motor positions cartridge
too high. place. in correct ___location.
2) Review 3) Check for lamp
cartridge function and drift.
handling Replace lamp if
instructions necessary.
3) Inspect 4) Run performance
cartridge test.
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
reference
channel light
path. (use
compressed
air)
6) Rerun
test
E302 The mean During Condensation, dirt or 1) Make 1) Blow out cartridge
Microalbumin / transmittance of Microalbumin / defect on cartridge sure cartridge area.
Creatinine Buffer the 10 buffer Creatinine test optical window; buffer is fully 2) Verify cartridge
Mean Absorbance readings sequence tab removed before inserted and holder integrity and
Too High (buf1...buf10) is inserting cartridge; or locked in motor positions cartridge
too low. buffer tab not pulled; place. in correct ___location.
or cartridge not 2) Review 3) Check for lamp
located correctly in cartridge function and drift.
holder; or optical handling Replace lamp if
problem; electronic instructions necessary.
drift 3) Inspect 4) Run performance
cartridge test.
window for
scratches or
finger prints
4) Run
optical test
cartridge.
Troubleshooting 5-13
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
5) Clean
cartridge
holder area,
dust blocking
reference
channel light
path. (use
compressed
air)
6) Rerun
test
E303 The coefficient of During Particulate 1) Make 1) Blow out cartridge
Microalbumin / variation (%CV) of Microalbumin / contamination in sure cartridge area.
Creatinine Buffer the 10 Creatinine test cartridge; or capillary is fully 2) Verify cartridge
Coefficient of transmittance sequence tube broken and inserted and holder integrity and
Variation Too readings taken of starch plug in assay locked in motor positions cartridge
High the buffer solution buffer; or high "short place. in correct ___location.
in the reagent term" instrument drift 2) Review 3) Check for lamp
cartridge is too cartridge function and drift.
high. handling Replace lamp if
instructions necessary.
3) Inspect 4) Run performance
cartridge test.
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
light path. (use
compressed
air)
6) Rerun
test
E304 Sample The mean During Particulate 1) Make 1) Blow out cartridge
Blank Mean transmittance of 1 Microalbumin / contamination in sure cartridge area.
Absorbance Too of the 3 sets of Creatinine test cartridge is fully 2) Verify cartridge
Low sample blank sequence inserted and holder integrity and
reading is too high locked in motor positions cartridge
place. in correct ___location.
2) Review 3) Check for lamp
cartridge function and drift.
handling Replace lamp if
instructions necessary.
3) Inspect 4) Run performance
cartridge test.
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
light path. (use
compressed
air)
6) Rerun
test
Troubleshooting 5-14
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
E305 Sample The mean During Urine sample is turbid 1) Make 1) Blow out cartridge
Blank Mean transmittance of Microalbumin / or highly colored. sure cartridge area.
Absorbance Too the "low sample Creatinine test is fully 2) Verify cartridge
High blank" is too low. sequence inserted and holder integrity and
locked in motor positions cartridge
place. in correct ___location.
2) Review 3) Check for lamp
cartridge function and drift.
handling Replace lamp if
instructions necessary.
3) Inspect 4) Run performance
cartridge test.
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
light path. (use
compressed
air)
6) Rerun
test
E306 %CV The %CV of the During Particulate 1) Make 1) Blow out cartridge
Sample Blank "low sample blank" Microalbumin / contamination in sure cartridge area.
Too High is too high. Creatinine test cartridge; or cartridge is fully 2) Verify cartridge
sequence use life expired; or inserted and holder integrity and
cartridge exposed to locked in motor positions cartridge
high temperature place. in correct ___location.
and/or humidity; or 2) Review 3) Check for lamp
capillary tube broken cartridge function and drift.
and starch plug in handling Replace lamp if
assay buffer; or high instructions necessary.
"short term" 3) Inspect 4) Run performance
instrument drift cartridge test.
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking l
light path. (use
compressed
air)
6) Rerun
test
E307 Mean The mean of the During Cartridge exposed to 1) Make Most likely a reagent
Microalbumin Microalbumin Microalbumin / high temperature sure cartridge issue
Absorbance Too transmittance Creatinine test and/or humidity is fully
Low readings is too sequence inserted and
high locked in
place.
2) Review
cartridge
handling
instructions
3) Inspect
cartridge
Troubleshooting 5-15
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
light path. (use
compressed
air)
6) Rerun
test
Mean The mean of the During Cartridge exposed to 1) Make Most likely a reagent
Microalbumin Microalbumin Microalbumin / high temperature sure cartridge issue
Absorbance Too transmittance Creatinine test and/or humidity is fully
High readings is too low. sequence inserted and
locked in
place.
2) Review
cartridge
handling
instructions
3) Inspect
cartridge
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
light path. (use
compressed
air)
6) Rerun
test
E309 %CV of The %CV of the During Particulate 1) Make Most likely a reagent
Microalbumin Microalbumin Microalbumin / contamination in sure cartridge issue
Transmittance transmittance Creatinine test cartridge; or high is fully
Too High readings is too sequence "short term" inserted and
high. instrument drift locked in
place.
2) Review
cartridge
handling
instructions
3) Inspect
cartridge
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
light path. (use
compressed
Troubleshooting 5-16
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
air)
6) Rerun
test
E310 Creatinine One of 26 During Cartridge exposed to 1) Make Most likely a reagent
Absorbance Too Creatinine Microalbumin / high temperature sure cartridge issue
Low absorbance Creatinine test and/or high humidity is fully
readings is too low. sequence inserted and
locked in
place.
2) Review
cartridge
handling
instructions
3) Inspect
cartridge
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Rerun
test
E311 Creatinine One of 26 During Cartridge exposed 1) Make Most likely a reagent
Absorbance Creatinine Microalbumin / to high temperature sure cartridge issue
Too High absorbance Creatinine test and/or high is fully
inserted and
readings is too sequence humidity
locked in
high. place.
2) Review
cartridge
handling
instructions
3) Inspect
cartridge
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Rerun
test
E313 Mean The Mean Squared During Particulate 1) Make Most likely a reagent
Squared Error of Error calculated Microalbumin / contamination in sure cartridge issue
Creatinine after the third order Creatinine test cartridge; or capillary is fully
Absorbance Too polynomial fit of sequence tube broken and inserted and
Large the Creatinine starch plug in assay locked in
absorbance buffer; or cartridge place.
readings is too exposed to high 2) Review
large. temperature and/or cartridge
high humidity handling
instructions
3) Inspect
cartridge
window for
scratches or
finger prints
4) Run
optical test
cartridge.
5) Clean
cartridge
holder area,
dust blocking
Troubleshooting 5-17
Corrective
Error or When Corrective action
Description Probable Cause action
Message Detected Service
Customer
reference
channel light
path. (use
compressed
air)
6) Rerun
test
E315 The Microalbumin After scanning a The instrument has 1) Run the 1) Run normalization
Microalbumin Normalization has Microalbumin / not been normalized Normalization cartridge
Normalization Not not been Creatinine test (Optical Filter Test cartridge. 2) Reset system from
Set performed or the cartridge S3); or the MNF has 2) Reboot factory setting mode
Microalbumin been corrupted; or the the instrument 3) Replace main PCB
Normalization computer's database 3) Rerun
Factor (MNF) has has been corrupted test
been corrupted.
E316 The Microalbumin When running the Instrument 1) Clean 1) Run normalization
Microalbumin Normalization Microalbumin normalization cartridge cartridge
Normalization Out Factor being Normalization procedure did not holder area, 2) Reset system from
of Range entered is outside (Optical Filter Test correctly enter the dust blocking factory setting mode
of the expected S3) normalization factor. reference 3) Replace main PCB
range. OR The normalization channel light
cartridge was not path. (use
correctly located in compressed
the cartridge holder. air)
OR The normalization 2) Re-run
scatter filter is the
contaminated, dirty, Normalization
misaligned, or cartridge.
missing. OR Optical 3) Reboot
or electronic the instrument
instrument defect. 4) Return
instrument for
service
E70 Internal The Internal printer When the printing to Bad/missing 1) Return 1) Reseat printer
Printer Error has failed due to a the internal printer connection or instrument for cables in connectors
hardware error electronic failure service 2) Replace printer
3) Replace Main PCB
E71 External The External When attempting to Bad/missing 2) Make 1) Test external
Printer Error printer has failed connect or print to connection or sure printer printer port. Replace
due to a hardware the external printer electronic failure power is on. main PCB if necessary
error 3) Reseat
printer cable
on Vantage
and at printer
4) Turn
printer off then
back on
5) Verify
that external
printer works
using PC.
6) Return
instrument for
service
5-18 Troubleshooting
Test Description
Touchscreen test Tests if the touchscreen is functioning properly.
Internal barcode test Tests if the internal barcode is functioning properly.
Internal printer test Tests if the internal printer is functioning properly.
Serial port test Tests if the serial port is connected properly.
USB port test Tests if the USB port is connected properly.
Lamp test Tests if the lamp is functioning properly.
Door test Tests if the door is opening and closing properly.
External barcode Tests if the external barcode reader is functioning properly.
reader test
External printer test Tests if the external printer is functioning properly.
External port test (if Tests if the external port is functioning properly.
supported)
Motor control test Tests if the motor control is functioning properly.
Measurement system Tests if the measurement are working properly.
test
Test Description
System audible emitter Tests if the system audible emitter is functioning properly.
test
ASTM port test Tests if the ASTM port is functioning properly.
Lamp test Tests if the lamp is functioning properly.
Motor sequence test Tests if the motor sequence is functioning properly.
Touchscreen test Tests if the touchscreen is functioning properly.
Internal and external Tests if the internal and external barcode reader is
barcode reader test functioning properly.
Troubleshooting 5-19
Test Description
Temperature test Tests if the temperature is in the correct range.
Motor position test Tests if the motor is in the correct position.
Sample and reference Tests if the sample and reference channel measurement is
channel measurement correct.
test
Touchscreen test
Use the touchscreen test screen to test if the touchscreen is functioning properly.
5-20 Troubbl
eshoo ting
Sound test
Use the speaker test screen to test the sound volume. The system emits an audible alert when
the users select the Test button. The default Sound setting is Medium.
For more information see the ASTM Interface Specification, Doc. No. 16566.
NOTE: The user can turn Monitoring on and off an unlimited number of times. The user can
turn on and off the lamp and heaters independently from Monitoring.
6 Maintenance
Maintenance Schedule
Weekly As Needed
Cleaning the Barcode Window Cleaning the Exterior
Cleaning the Exterior Changing the Air Filter
Quarterly Removing and cleaning the cartridge
spring
Removing and cleaning the Cleaning the Barcode Window
cartridge spring
Changing the Air Filter Running the Optical Test
Running the Optical Test Calibrating the Touchscreen
Replacing the Fuse
WARNING: Turn the power off and disconnect the power cord before cleaning the
barcode window.
Materials required:
• lint-free cloth
• water or ethanol
6-2 Maintenance
1. Clean the barcode window with a lint-free cloth dampened with water or ethanol.
WARNING: Turn the power off and disconnect the power cord before cleaning the
exterior of the system.
CAUTION: Do not allow liquid to drip into system. If liquid drips into the system, you
can damage the optics.
Materials required:
• lint-free cloth
• water or ethanol
1. Clean the exterior with a lint-free cloth dampened with water or ethanol.
2. Connect the power cord after the exterior is clean and dry.
NOTE: If you want to disinfect the exterior of the system, expose the surface to 0.5% sodium
hypochlorite for 10 minutes. Remove any liquid blood on the system before disinfection.
CAUTION: Do not use any other type of solvent, oil, grease, or silicone spray on any
part of the system.
Maintenance 6-3
WARNING: Turn the power off and disconnect the power cord before cleaning the
cartridge compartment.
CAUTION: Do not allow liquid to drip into system. If liquid drips into the system, you
can damage the optics.
Materials required:
• lint-free cloth
• water or ethanol
• paper clip or similar device
• mild detergent
• sponge swab
1. Open the cartridge compartment door as far as possible.
2. Wipe the inside surface of the compartment door and surfaces on both sides of the
compartment using a lint-free cloth dampened with water or ethanol.
3. Dry the surface using a clean, dry, lint-free cloth.
4. Locate the cartridge return spring inside the cartridge holder.
6-4 Maintenance
5. Insert the tip of a straightened paper clip (or other like device) into the top hole on the
spring.
1 Top Hole
2 Return Spring
3 Bottom Hole
6. Gently pull the metal end towards the center of the cartridge compartment to release one
side of the spring from the cartridge holder.
7. Repeat step 6 to release the other side of the spring from the cartridge holder.
8. Pull the cartridge return spring completely out of the system.
9. Clean the cartridge return spring using any of the following items:
• warm solution of mild detergent and water
• lint-free cloth dampened in water or ethanol
Maintenance 6-5
NOTE: Make sure the leaf springs are not bent or damaged while cleaning. Damaged leaf
springs do not function properly.
1 Return Springs
10. Dry the cartridge return spring with a clean, lint-free cloth.
11. Using a clean, dry, sponge swab (provided in the Cleaning Kit), remove spilled liquid
from the cartridge holder.
CAUTION
Do not use a cotton swab. Cotton fibers that are left on the surface
can interfere with the system’s optical systems.
6-6 Maintenance
12. Rotate the cartridge holder with the compartment door partially closed to locate and
remove any additional liquid.
CAUTION: Do not allow liquid to drip off of the sponge swab into the system. If liquid
drips into the system, you can damage the optics.
1 Filter Cover
DRAFT
6-8 Maintenance
NOTE: The cartridge is designed to fit one way into the system.
CAUTION: Do not touch the printer without observing precautions for handling
electrostatic sensitive devices. A risk of electrostatic discharge to the system exists
when touching the printer.
c. Carefully pull paper through the printer in its normal direction of travel.
1 Printer Cover
2 Paper Feeder
3 Paper Cover
11. Push the paper gently under the roller at the back of the printer.
CAUTION: Set the power switch to OFF. Disconnect the power cord from the wall
outlet. Disconnect the power cord from the system.
1 Fuse
2 Fuse
3 Fuse Block
7 Diagrams
Overview
The main PCB provides all of the hardware for the following areas:
• user interface processing
• realtime processing logic
• data acquisition
• control and utility functions
The main PCB interfaces with the measurement module, LCD touchscreen display, thermal
printer, and the power supply functions.
Circuit Descriptions
The following sections explain the circuit descriptions.
Microprocessor
The main DCA Vantage instrument uses a 32 bit microprocessor that operates at an internal
clock frequency of 200 megahertz. The external synchronous memory bus operates at 100
megahertz, and it includes a realtime clock and requires battery power to maintain time when
the unit is disconnected from AC power.
Nonvolatile memory
DCA Vantage includes 128 megabytes of nonvolatile memory.
Volatile memory
DCA Vantage includes 128 megabytes of standard SDRAM volatile memory.
• current signals
The digital signals have a minimum resolution of 17 bits.
Microalbumin/Creatinine Assay
The microalbumin/creatinine cartridge assay is a quantitative measure of the concentration of
albumin in urine.To normalize measured albumin for urine volume variation, creatinine is
measured in the same cartridge and the ratio of 2 measurements is reported in addition to the
individual concentrations of albumin and creatinine. A reporting units option is available to
select between mg/dL and mmol/L units for creatinine concentration and ratio units. Ratio
units are not reported, except in the recall patient results menu.
The reagent cartridge contains all the reagents necessary for the assay. The assay requires 40
microliters of urine, pipetted into the cartridge using a dedicated capillary holder.
The microalbumin/creatinine calibration card includes a barcode on each side of the card that
provides data specific to the lot number of each microalbumin/creatinine reagent kit. You
have to scan both barcodes to enter calibration data into the instrument (Malb/C and Malb-2).
Diagrams 7-5
5. Buffer solutions flows to the read area of the cartridge and a blank reading is taken.
7. The cartridge rotates to the read position where sample blank reading is taken.
9. The cartridge rotates to the read position where microalbumin reading is taken.
11. The cartridge rotates to the read position where creatinine reading is taken.
HbA1c Assay
In humans, red blood cells (erythrocytes) normally have a life span of approximately 120
days. During this period, certain components of the protein hemoglobin (within the red cells)
slowly combine with the glucose in the blood in an irreversible process referred to as
glycosylation. The higher the glucose and the longer it is elevated, the more it is retained by
the hemoglobin. Quantifying this glucose component can determine the degree of glucose
control attained by a person with diabetes.
Hemoglobin consists of several subfractions including Hemoglobin A (HbA), which
comprises 97% of the total, HbA2 is 2.5% and HbF is 0.5%. HbA is composed of glycated
subfractions collectively referred to as HbA1. Subfractions of HbA1 are HbA1a, HbA1b, and
HbA1c. HbA1c comprises 80% of HbA1 and is the subfraction of interest in terms of indexing
the degree of blood glucose control.
The quantification of HbA1c is expressed in percent (%) of total hemoglobin:
The elevation in the diabetic of HbA1c was first described in the late 1960s, but testing did
not progress into clinical use until the 1980s. Clinical acceptance of the test coincided with
the clinical philosophy that the diabetic person’s glucose pattern should mimic (within
practical limits) the of the non-diabetic’s. Since the glycosylation of hemoglobin is slow and
irreversible, the HbA1c assay provides a long term diabetic metabolic control or assessment
of mean blood glucose. This allows physicians to know at a glance the degree of metabolic
control their patients have experienced over the last 60 to 120 days.
The DCA Vantage test determines both total hemoglobin and HbA1c using 1 µL blood
sample.
Capillary holder
The capillary holder is designed to obtain the test sample and insert into the reagent test
cartridge. The holder consists of a 1 µL heparinized capillary positioned at the end of a
molded plastic holder. The holder is used to obtain a 1 µL sample of blood from a finger
prick. The holder is designed to insert and lock into the test cartridge by a locking tab. When
inserted, the capillary tip comes in contact with the liquid buffer during cartridge rotation.
NOTE: Testing must be done within 5 minutes from the time blood is picked up in the
capillary holder.
7-12 Diagrams
The capillary holder contains an absorbent sponge within the plastic holder. The sponge is
used to absorb test fluid at the end of the testing sequence. The cartridge is rotated to an
absorbent picking up position where time is allowed for fluid absorption.
Barcode label
The cartridge barcode label is designed to pass through the instrument barcode reader track.
The label must face the code reader (black dot) and can pass though the reader in either
direction.The barcode label on the test cartridge contains the HbA1c test type identification
and the reagent lot number. The cartridge barcode initiates the HbA1c processing sequence
and retrieves calibration data from the instrument memory.
Diagrams 7-13
Optical window
The cartridge optical window aligns to the cartridge holder sample channel optical aperture.
Inserting and snapping the cartridge into the instrument cartridge holder automatically aligns
the window to the holder aperture. The window is then placed in the instrument read optics
path by holder rotation.
Oxidant
Oxidant oxidizes hemoglobin to methemoglobin. Picked up by buffer and reconstituted with
sample during the first mixing phase. Following a mix and 2 minute reading incubation
phase, a total hemoglobin reading is taken.
Antibody latex
Antibody coated latex particles.
Agglutinator reagent
Picked up and reconstituted by cartridge rotation and mixed.
2. After the barcode scan, the cartridge is inserted into the instrument.
3. The buffer release tab is removed and discarded
4. The buffer solution is released into the read area of the cartridge where a blank reading is
taken.
Diagrams 7-15
6. The cartridge rotated to the read position where a hemoglobin reading is taken.
8. The cartridge rotates to the read position where an HbA1c reading is taken.
9. When the assay sequence is complete, the cartridge rotates to a position allowing the
liquid to flow into the capillary holder absorbent pad.
Calibration Card
Each reagent kit contains a calibration card with a barcode label which is designed to pass
through the instrument barcode reader track. The calibration barcode identifies the test type,
lot number, expiration date, and calibration constants. The calibration barcode initiates
calibration data memory storage. Calibration storage data is provided for the 2 most recently
scanned lots of reagents and for up to 8 different reagent types.
Calibration constants are determined for each hemoglobin reagent lot. The constant values
are then encoded in the barcode on the calibration card.
Each normal or abnormal control contains a control card with a barcode label. The control
card identifies the test type, control lot number, level identifier, date of manufacture,
expiration range, and checksum. The checksum verifies the code integrity.
Motor timing
The cartridge holder loading and default reading positions are the same and corresponding to
the motor home position, The Home/Index sensor input provides transitions generated from
the holder post/slots which do the following:
1. Locate the home slot during homing.
2. Area accumulated during stepping sequence for confirming holder ___location.
3. Detect rotation limits at the fully CW and CCW positions.
During homing a final home offset (within 4 half-steps) is determined using optical channel
readings. Readings are taken through the holder home viewing apertures for the sample and
reference channels.The offset value allows for tolerance and assembly variations between the
true optical centerline and the home sensor viewing pattern.
During cartridge rotation sequences, occurring transitions are counted by the home/transition
sensor and compared to a memory stored value required for that rotation. If the two values do
not match, an error code 12 is set and the motor inhibited. If the transition value exceeds the
CW or CCW rotation limits, error code 12 is set.
Home/Index sensor
The Home/Index sensor is read to determine the which of the following is the current status:
• If blocked (H) the rotation is CCW
• If clear (L) the rotation is CW
The motor is half stepped in the direction determined and the sensor is read each step until
one of the following events occurs:
• 10 sequential clear readings (L) occur, if the motion is CCW direction, the motion is
changed to CW direction until 12 sequential block readings (H) occur.
• 12 blocked readings (H) occur, If the motion is in the CCW direction, the motion
continues until the home edge is detected. If the motion is in the CW direction, the
motion is changed to the CCW direction and continued until the home edge is found.
Diagrams 7-19
Following the CCW 12 blocked readings, the home edge is located with half stepping motion
when 3 high to low transitions occur. The home edge is indicated by the third transition.
Slot/Post pattern
The slot patterns are detailed in figures above. The full step angular displacement to the
various cartridge positions are given. The actual testing step sequence depends on the test,
for example, HbA1c.The home and read locations correspond directly to the indicated
positions. The true angular blood sample and reagent pickup with mixing do not correspond
to the actual angular post/slot position for that activity. Sample and reagent pickup is
achieved by a washing action through the cartridge rotational action. Program controlled
rotational limits achieve pickup, assure thorough mixing and avoid fluid loss by over
rotation.
Diagrams 7-21
Barcode symbology
The barcode reader is configured for interleaved 2 of 5 symbology (numerical). The barcode
must have an even number of characters including a checksum and must be bracketed by a
start and stop pattern. The maximum numbers of characters is 40. In this encoding, characters
are paired and represented by 2 wide elements and 3 narrow elements. The 5 elements are
represented by bars for the most significant digit and spaces for the least significant. This
allows high encoding density ranging up to 18 characters per inch. The encoded binary
values by position include the following:
The stepper motor is secured to the motor mount with the cartridge holder assembly mounted
with the cartridge holder assembly mounted on the stepper motor shaft. The stepper motor
rotates the cartridge holder in both directions from a cartridge loading position.
Figure 7-17 Cartridge holder assembly
Stepper motor
The stepper motor is a 200 step (1.8 degrees per full step) device mechanically restricted to
340 degrees. Maximum rotation rate is 167 steps/seconds (50 rpm) program optimized for
each reagent test type (HbA1c). Optical read positions are located at full step positions. The
motor is normally operated in a half step mode during rotation. Motor timing inaccuracy is
less than 1% per half step.
All motor movements are performed with the +12VDC drive. Optical readings are taken with
one motor winding energized to provide a stable and repeatable motor state. At a motor stop
position, a 5 ms setting time is allowed between the last step of motion and switching to a
standby +5VDC value. A 200 ms settling time is then allowed before the start of a
transmission reading.
Motor rotation is inhibited when the cartridge access door is opened. An optical door sensor
monitors the door ___location by generating the signal DOOR. This signal is coupled to the
controller PCB gate array U21.
Diagrams 7-23
When the instrument is in the standby state, all motor current is disabled from the motor
drive circuit.
The cartridge is referenced vertically and front-back by a removable U shaped return spring.
A rear leaf on the return spring positions the cartridge forward against the two limit edge rails
described above. A bottom leaf positions the cartridge locking tab upward in the fixed plate
locking tab slot.
The 4 pressure springs are positioned and secured by upper cartridge holder tabs and by 2
bottom cover braces. The bottom cover snaps to the upper holder and is removable for
service.
Slot/Post Ring
The cartridge holder assembly contains a slot/post ring which indexes to the home/index
sensor. This indexing provides positioning feedback during holder rotation to specific
locations. The ring/post locations reference to the read apertures. A cartridge sample read
aperture (sample channel) is located at the bottom of the holder. The reference channel read
aperture and sample reference channel Home/Air reading apertures are located in the holder
cover.
These components are designed to electronically achieve the required reagent cartridge
reactivity temperature within an ambient temperature range of 15 - 32C°. Heater plate to
cartridge flatness is designed for an air gap of 0.005 inches. The thermistors are selected to
operating characteristics.
The above components are electrically connected to the controller board through a 12 inch
spiral flex cable. When the holder rotates the lead winds on the molded cable wrap portion of
the holder.
Locking yoke
A holder locking yoke locks the holder assembly at the cartridge load home position when
the cartridge door is opened. Opening the door allows the spring loaded yoke to engage a
holder locking track slot. Closing the door forces the locking yoke downward freeing the
holder to rotate on the motor shaft.
Home/Index sensor
The home/index sensor is used to sense critical cartridge holder rotary positions. The sensor
is a motor mount located to detect holder posts/slots edges at specific rotary positions. The
sensor Home/Index signal line inputs to the controller board gate array. The Home/Index
signal provides the following:
• A homing signal during cartridge holder homing
• Confirming signals during cartridge rotation to sample and reagent pickup locations, and
during mixing
When the motor is moved to a specific locations, the index signal changes state indicating
that positions.
The operating system includes the following 3 sensor failure error codes:
• E10
• E11
• E12
For more information see Troubleshooting .
7-26 Diagrams
Instrument optics
CAUTION: All DCA Systems instruments read blank corrected neutral density
and scatter filter transmissions to within the tolerances established by a
nominal or standard instrument. This tolerance testing is provided by the
individual filter test module sets and performance test software program.
The instrument optical scatter reading system is diffused transmission n design optimized for
the highest resolution possible. The system is integrated into the motor mount assembly and
consists of the source optics, cartridge holder apertures, reagent cartridge optics and detector
optics. The sample area optics include both the cartridge holder apertures and the reagent
cartridge optical window. The sample area of the optical system is stepper motor rotated
between the source and detector optics.
To compensate for lamp intensity variations during the testing period, a sample
chance/reference channel design is used. A transmission ratio between the 2 channels
compensates for variations. The sample channel measures light transmittance through the test
cartridge (reagent sample) and reference air reading aperture. The reference channel
measures reference light levels through air read apertures. Light source variation affect both
channels proportionally. The ratio between the 2 channels is directly proportional to the light
available without regard to light source variations.
Source optics
The source optics are designed to create a controlled beam of light to illuminate 2 optical
channels. The source optics consist of a incandescent halogen lamp, 0.025 inch collimating
aperture, spacer, and lens. The lamp and source components are mounted in a metal holder to
maintain illumination alignment under thermal conditions. The nominal lamp tip on axis
distance from the aperture is 1.5 mm (0.06 inch). The lamp is positioned and mounted by
dowel pins to minimize heat transfer and lamp heat temperature shift. A flat screw secures
the lamp and aids in temperature shift. A 531 nm interference filter mounts into the source
housing cap and is secured by 1 or 2 foam O rings as required. The source optics assembly
locates into a molded area on the motor mount. Instrument mounting is self aligned by guides
on the mount which locate the cap, and by a fixed vertical pin guide into a base mount slot.
The vertical dowel pin establishes the assembly positioning the motor mount.
Light from the collimation aperture strikes the lens and is formed into a controlled beam. The
beam passes through the 531 nm filter. 531 nm is the active wavelength of chemistries tested
on the DCA System. The band pass output of the filter is then formed into the sample and
reference channels by source exit apertures molded into motor mount. These two channels
pass through the sample area optics (cartridge/holder) to the detector optics. Each channel is
independently illuminated and isolated during readings.
Figure 7-21 Source optics assembly
Cartridge/Holder optics
The cartridge holder optics consists of the following:
• read apertures (sample and reference channels)
• home/air apertures (sample and reference channels)
• dark reading locations
The apertures are designed to align with the source and detector optics as the cartridge is
rotated. Reagent cartridge readings occur at the read apertures. the reference channel aperture
is an air reading. The home/air apertures are used during the cartridge holder homing
sequence and for air readings during sample testing. Air readings establish the air ratio. Dark
or offset readings are taken at a cartridge ___location between the read apertures and home/air
apertures. This ___location blocks light source transmission during lamp on times providing the
electronic offset value.
The reagent cartridge optical window is designed for the following:
• 5.1 mm square light entry window
• 2 mm read area (cylinder)
• 1 cm optical path length (standard cover)
The cartridge holder and cartridge window design accommodates the 2 mm detector optic
read cylinder. The 1 cm cartridge optical path length is maintained at +/- 0.11 mm (0.0043
inch) maximum to minimize cartridge to cartridge path length variations. The cartridge
optical window is specified to be free of chips, scratches, oil, flash, or any marks. Each
window surface is specified or 91% transmission with no defects.
7-28 Diagrams
Detector optics
The detector optics consists of serrated collimating tubes for each channel which couple to
photodiode/preamplifier detectors. The collimating tubes are fixed in the motor mount and
align to the lamp source exit apertures. The selected length of the tubes maximize
transmittance to the photodiodes. The internal tube serrations establish the 2mm read area
cylinder with a 5 degree read angle cone for minimizing internal stray light. The read
cylinder optimizes scatter resolution and allows a usable scatter range of 10% to 100%
The 2mm read area cylinder establishes a 0.5 mm read movement across the optic window
for each motor half step (0.9 degree). The actual read ___location is established in the center of
the read aperature.
A.2 Installation
The application is installed from the CD. Run Setup.exe
A.2.1 Procedure
1. Verify that Microsoft Access 2002 is installed on the PC that is being used. If it is not
installed obtain a copy and install it.
2. Verify that Microsoft .Net framework 1.1is installed on the PC. If it is not then insert the
CD contain the test software and find the file “dotnetfx.exe”. Run this program.
3. The test software application is installed from the CD by Run “Setup.exe”. This will take
you through the installation process. It will add an icon to the Desktop. Double-clicking
this icon will run the application.
4. Any updates to the application will require the existing database to be renamed prior to
uninstalling the application. Note: If this process is not followed the data in the
existing Database will be overwritten.
Appendix A A-3
Appendix A A-4
Appendix A A-5
Appendix A A-6
5. Under the Configuration Details tab, the initial installation will point to the Access
database. Ensure that this is always adhered to in the Service environment.
6. The “list installed printer” will detail printers available, and show the default printer.
7. Printer configuration is default to “off”. If the printer configuration is set to “on” it
will automatically print a report at the end of the test, if set as default (“off”), an
option to print is available at the end of the test.
8. The default serial port for instrument communication is “1”, if this port is not
available, a message will be displayed, and alternative ports can be selected via the
drop-down box.
9. Save new configurations before exiting.
Appendix A A-9
10. The Specification Details tab contains the specification limit values for DCA Vantage
and previous models. Note these are identical currently, so this feature is there for
future-proofing. It also refers back to the source of these values and communication
settings, again common for all instruments. These values should not be changed in-
less directed to via service bulletin!
Appendix A A-10
Appendix A A-11
Appendix A A-12
Appendix A A-13
11. The Calibration Details tab allows service engineers to add new filter set values and
update as necessary. These assigned values should be taken from the A1CSTD file
from Mishawaka.
12. Use the “Add” button to create new details. Enter the filter set number in the box (3
digits, including any leading zeros) in the right hand corner and assigned values
(X.XXX) and calibration details. “Save” details before exiting. Double clicking on
the number of an existing entry will reveal the assigned values and calibration details
for that set, which can then be updated. “Save” details before exiting.
13. Press “Cancel” to exit the “Edit Configuration” panel.
Appendix A A-14
1. The 1st test instructs the user to connect the serial cable, and that the instrument is
ready for test. Press “Pass” to continue, “Fail to abort”.
2. The 2nd test instructs the user to enter the serial number into the DCA. Press “Pass” to
continue, “Fail to abort”.
Appendix A A-15
3. The 3rd test instructs the user to check the operation of the speaker. Press “Pass” to
continue, “Fail to abort”.
4. The 4th test instructs the user to check the operation of the touchscreen. Press “Pass”
to continue, “Fail to abort”.
5. The 5th test instructs the user to check the operation of the internal printer. Press
“Pass” to continue, “Fail to abort”.
6. The 6th test instructs the user to check the operation of the fan. Press “Pass” to
continue, “Fail to abort”.
7. The 7th test instructs the user to look for any adverse conditions that are not covered
by the previous tests. Press “Pass” to continue, “Fail to abort”. Pressing “Fail” will
allow the user to record the observations.
1. To run these tests carefully follow the on-screen instructions on the PC. When
installing filters into the instrument, carefully follow the instructions on the DCA.
Once “Start test” has been pressed, the test has to run to completion. A countdown
timer displays the maximum time left to complete the test. Note that when the test
completes and the data transfers to the PC, the test cycle progresses regardless of the
countdown timer. A total of 3 opportunities to test each filter are available to allow
any cleaning of filters, etc. A “Cancel test run” button is available at the completion
of any individual filter test; data is not saved to the database in this instance.
2. After the filter testing has been completed, the instrument must be reset, following the
on-screen instructions on the PC. Press “Pass” to continue, “Fail to abort”.
3. The report details for the test are then presented on-screen.
Appendix A A-16
The following notes give a brief explanation of the database structure for guidance in the
event that the Service group needs to write queries to the database.
Within the database structure the starting point in extracting data is the tblTestRuns. From
this table the serial number of the instrument can be found and this will correspond to an “id”
field. Using this “id number” in the tblFilterTests, the “filter id’s” for that test run can be
found. Using those “filter id’s” in the tblFilters the pass/fail status can be found. Individual
filter test data can then be extracted from the tables listed below. These contain the individual
test values, limits and pass/fail status.
Overview
This section defines the data codes, transmission protocols, error recovery, and message
content of communication between the DCA Vantage system (the instrument) and a
laboratory computer (the host).
The DCA Vantage system is capable of supporting a serial connection and/or an Ethernet
connection between instrument and host. The serial interface is mechanically and electrically
compatible with the EIA-232-D Standard. The Ethernet interface is established through the
instruments RJ45 internet connection.
The DCA Vantage system supports the ASTM Specification E1381, Low-Level Protocol to
Transfer Messages between Clinical Laboratory Instruments and Computer Systems. The
system also supports the ASTM Specification E1394, Standard Specification for Transferring
Information between Clinical Instruments and Computer Systems.
The following are ASTM messages transferred from instrument to host:
• HbA1c reagent test result records
• Microalbumin/Creatinine reagent result records
• Filter Test Cartridge result records
• Control Test result records
The following are ASTM messages transferred from host to instrument:
• Force Control remote commands
• Remote Lockout remote commands
For more information about setting up LIS specifications, see the DCA Vantage Operator’s
Guide.
External Documents
Use the documents in the following table as a reference.
LIS Terminology
The following table defines terms used for LIS interface specification:
Term Description
Establishment This is the phase in the ASTM 1381 protocol that determines the
Phase direction of information flow between instrument and host. The sender
requests to send by issuing an <ENQ>. The receiver accepts the request
by issuing an <ACK>.
Filter Test A test used to verify the optical performance of the instrument. These are
also known as Optical tests.
Frame A frame is a subdivision of an ASTM 1381 message. It is used to allow
periodic communication between devices. A message containing 240
characters or less is sent in a single frame. Larger messages are broken up
into multiple frames.
Host Any laboratory computer system as defined by the ASTM 1381 standard.
These computer systems typically collect reports from the instrument
and/or send remote commands to the instrument.
Instrument The DCA Vantage system is a Clinical Laboratory Instrument as defined
by the ASTM 1381 protocol.
Message A message is a collection of related information. It is the textual body of
information that is transmitted between instrument and host during one
session. Message can consist of many different types of records.
Session A session is a total unit of communication activity between instrument
and host as defined by the ASTM 1381 protocol. The session begins with
the establishment phase and ends with the termination phase.
Termination This is the phase in the ASTM 1381 protocol during which the sender
Phase relinquishes its control of the communication link by issuing an <EOT>.
After the termination phase, the communication link is returned to a
neutral state.
Transfer Phase This is the phase in the ASTM 1381 protocol during which the sender
sends messages to the receiver.
Physical layer
The Physical Layer refers to the point-to-point connection between the DCA Vantage system
(the instrument) and a laboratory computer system (the host).
3-159
Serial connection
The serial transmission between the instrument and the host is established through the
instrument's serial port using a serial cable. The mechanical and electrical connection for
serial binary data bit transmission between instrument and computer system is described in
the serial physical layer. The DCA Vantage system implements and complies with serial
physical layer defined in the ASTM 1381 standard. For additional information about this
layer, please refer to the standard.
Ethernet connection
TCP/IP transmission between the instrument and the host is established through the
instrument's Ethernet network adapter using a RJ45 connector. The DCA Vantage system
only supports the instrument connecting to the host. This means that the host needs to listen
for connections on a TCP/IP port and needs to accept the connection when the instrument
tries to connect.
"H" is field 1, "\^&" is field 2, "DCA Vantage^1.0.0.0" is field 5, "P" is field 12, and
"20060823235959" is field 14.
NOTE: In ASTM Records, the first field is always the record type.
Field 5 contains "DCA Vantage^1.0.0.0". This field contains 2 values separated by the
component delimiter.
Record types
An ASTM message consists of a hierarchy of records of various types. In ASTM Records,
the first field is always the record type. The following record types are used by the DCA
Vantage system:
• Message Header Record (H)
• Patient Record (P)
• Order Record (O)
• Result Record (R)
• Comment Record (C)
• Manufacturing Record (M)
• Message Terminator Record (L)
3-162
Example:
L|1|N<CR>
Example:
P|1|987654|||Doe^Jane<CR>
Result records are only transmitted from instrument to host. A result record can never appear
without a preceding patient record.
There are seven Manufacturing Record Types used by the DCA Vantage system. They are:
• Remote Command request record (host to instrument)
• Remote Command response record (instrument to host)
• Begin Data record - General information about a test (Service mode 1 or 2)
• Raw Data record - Raw data readings for a test (Service mode 1 or 2)
• Microalbumin Normalization Factor record (Service mode 1 or 2)(Microalbumin/
Creatinine test only)
• Temperature Readings record (Service mode 2 only)
• Total Hemoglobin record (Service mode 2 & HbA1c tests only)
3-167
Field Definition
MF1 "C" Indicator for Remote Command Request
MF2 MF2 Remote command request type:
"h" Force High Microalbumin/Creatinine Control
"l" Force Low Microalbumin/Creatinine Control
"a" Force Abnormal HbA1c Control
"n" Force Normal HbA1c Control
"s" Set Patient Lockout
"c" Clear Patient Lockout
MF3 The token provided by the host to identify the remote command request.
Field Definition
MF1 D Indicator for Remote Command Response
MF2 Remote command response:
"invalid login" User did not specify proper password in the ASTM
Header record
"Invalid request" Remote command type requested was invalid
"pending" System is Busy. System will force the control once
ready.
"started" System started the requested operation.
"completed" System completed the requested operation.
"timeout" Timeout waiting for the operation to start.
"error" Error occurred while processing the requested
operation.
MF3 The token provided by the host to identify the remote command request.
3-168
Field Definition
MF1 R - Indicator for Result Information Record
MF2 Tag - Text label - "air1", "dark1", etc…
MF3 Test time (10th of seconds from start of test)
MF4 Motor position (steps from optical home)
MF5 Mean sample channel reading (in millivolts)
MF6 Mean reference channel reading (in millivolts)
MF7 Mean S/R ratio (sample/reference)
MF8 C.V. of sample channel reading (in %)(coefficient of variance)
MF9 C.V. of reference channel reading (in %)
MF10 C.V. of S/R ratio
MF11 Transmittance (mean sample position / mean air position)
3-170
Temperature record
The Temperature record contains the following data:
Field Definition
MF1 R - Indicator for Result Information Record
MF2 HTRL or HTRR - Response Tag (heater left & heater
right)
MF3 Average temperature during test
MF4 Maximum temperature during test
MF5 Minimum temperature during test
MCRF record
The Microalbumin Normalization Factor record contains the following data:
Field Definition
MF1 R - Indicator for Result Information Record
MF2 MCRF - Response Tag
MF3 Normalization factor
THB record
The Total Hemoglobin record contains the following data:
Field Definition
MF1 R - Indicator for Result Information Record
MF2 THB - Response Tag
MF3 Total Hemoglobin value
MF4 Measurements units
3-171
Field Definition
MF1 C
MF2 1,2,3...
MF3 I Clinical instrument system
MF4 1 to many record specific text strings separated by a
component delimiter
MF5 G General/Free text comment
Before HbA1c results, transmit User Correction Slope, Offset, and Algorithm
C|1|I|1.000^0.0 %^default|G<CR>
Modes of Operation
The DCA Vantage system can operate in one of three modes, service mode disabled, service
mode 1, and service mode 2. Test result messages sent from the instrument to the host differ
based on the mode of operation. The following sections give an overview of these
differences.
3-172
Service mode 1
If service mode 1 is enabled, the system creates a Manufacturing Mode 1 message for the
current test in progress. Manufacturing Mode 1 messages have the following notable
characteristics:
• Message Header Processing ID = D for Debug
• User ID is not sent with the Result Record
• Manufacturing messages are sent to the host with the results
• Manufacturing messages contain:
• Manufacturing Mode Header
• Raw Data Readings of optics channels
• Microalbumin Normalization Factor if Microalbumin/Creatinine test
• No Comment records are transmitted in Service Mode 1
Service mode 2
If service mode 2 is enabled, the system creates a Manufacturing Mode 2 message for the
current test in progress. Manufacturing Mode 2 messages have the following notable
characteristics:
• Message Header Processing ID = D for Debug
• User ID is sent with the result record if entered
• Comment records are transmitted if the user entered text for patient demographics.
• Manufacturing messages are sent to the host with the results
• Manufacturing mode 2 message contain:
• All Manufacturing records for mode 1
• Temperature Readings
• Hemoglobin information if HbA1c test
Transmit Use-Cases
This section describes use-cases that involve transmission of ASTM message between the
DCA Vantage system and the LIS host. It also describes the messages transmitted during
those use-cases.
3-173
R|1|^^^Alb|5.0|mg/L||<|||||20061002183420<CR>
R|2|^^^Crt|15.0|mg/dL||<|||||20061002183420<CR>
R|3|^^^Ratio|---|mg/g|||||||20061002183420<CR>
L|1|N<CR>
L|1|N<CR>
Remote commands